AIC-2500 Troubleshooting Checklist

If you find that your table height is not working properly or is stuck in the down position:


Step 1-Double check to see if the table is plugged in.


Step 2- Check to see if the table is in the “On” Position, the “I” depressed.
This is to assure that the battery is not drained due to no line power. This is caused by the table shipment or if it is moved throughout the facility. The battery backup has an expected working life of a fully charged battery which is approximately 5 hours at 10% Duty Cycle.


Step3-Once steps 1 and 2 have been completed, lower the table until it is about a ½” away from the bottom. Then hold the down button on the hand control for 15-30 seconds.


Step 4-If the above steps still do not rectify the problem, you will then need to synchronize the columns.
You were given with your table shipment a box of extra fuses, and an additional hand controller that is different from the one that comes standard with your table. That controller is the one used for the synchronization process. This malfunction is caused by the battery draining to the point where the memory functions have been deleted, which is why it is very important to always have the batteries fully charged. This is done by keeping the table plugged in as often as possible, and having the table turned on with the “I” depressed.


IMPORTANT: Prior to synchronizing the columns, please call Tower Medical Systems, Ltd. for a walk through of the procedure. This process will erase all memory settings and may cause your table to not function properly if it is not done correctly.

 